Brand Name : JUGE
Model Number : AG6520A
Place of Origin : Nanjing, China
Certification : IOS9001, CE
MOQ : 1 Unit
Price : USD300-600
Packaging Details : Neutral carton
Delivery Time : 5-8 working days
Payment Terms : Online order,T/T
Supply Ability : 100 Unit/Week
Ion : Nitrate Ion Electrode
Range : 0.4–62,000 mg/L(2.5-11pH)
Temperature Range : 0–50°C
Pressure Resistance : Not pressure-resistant
Housing Material : Epoxy (EP)
Connection Thread : PG13.5
Nitrate Ion Sensor ISE Probe
>>Technical Parameter
Type: PVC membrane ion-selective electrode
Measurement Range: 0.4–62,000 mg/L
pH Range: 2.5–11 pH
Temperature Range: 0–50°C
Pressure Resistance: Not pressure-resistant
Temperature Sensor: None
Housing Material: EP
Connection Thread: PG13.5
Cable Length: 5 m or custom
Cable Terminal: Tinned or custom
>>Overview
The ion-selective electrode (ISE), also known as a membrane electrode, is an electrode with a sensitive membrane that generates a membrane potential based on ion exchange or diffusion. It primarily consists of an ion-selective membrane, an internal reference electrode (typically Ag/AgCl), and an internal reference solution (composed of chloride and strong electrolyte solutions of the target ion). The ISE measures the electrical potential of a specific ion in a solution, which is determined relative to a reference electrode with a constant potential. The potential difference between the two electrodes depends on the activity of the specific ion in the solution. This activity is related to the ion’s concentration, enabling quantitative analysis.
>>Feature
>>Pre-Measurement Preparation
>>Calibration (Slope Determination)
Connect the electrode to the meter.
Pour 100 mL of deionized water into a 150 mL beaker and stir thoroughly. Set the meter to mV mode.
Blot the rinsed electrode dry and immerse it in the prepared deionized water.
Add 1 mL of 1000 ppm standard solution to the beaker, stir thoroughly, and record the stable reading (E₁).
Add 10 mL of the same standard solution to the beaker, stir thoroughly, and record the stable reading (E₂).
The slope at 25°C is calculated as E₂ – E₁. Compare the result with the specifications to validate electrode performance.
>>Troubleshooting
If the slope deviates from the specified range:
>>Storage
![]() |
PG13.5 Thread Nitrate Ion Sensor ISE Probe With 5m Or Custom Cable Length Images |